Model Number 5001102-AS |
Device Problems
Leak/Splash (1354); Nonstandard Device (1420); Failure to Prime (1492)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/28/2015 |
Event Type
Injury
|
Manufacturer Narrative
|
There are no devices of this lot remaining in inventory for analysis.Quest medical, inc.Has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Quest medical, inc.Defers to the patient's physician regarding medical history.
|
|
Event Description
|
The hospital perfusionist reported an issue encountered with the mps delivery set during use.She reported that the delivery set was observed to be leaking during priming, prior to the start of the surgery.The delivery set was changed out prior to use during the surgery.There were no patient complications reported as a result of the alleged event.The device was returned to the manufacturer for analysis.
|
|
Manufacturer Narrative
|
The returned complaint sample pump cassette was attached with air source and dunk tested in water to identified leak.The leaking was identified below the blood inlet bushing and delamination of weld seam area confirming complaint condition.Dhr records were reviewed for the same lot and no rejected devices or manufacturing issues were documented similar to this reported complaint condition.
|
|
Search Alerts/Recalls
|