Catalog Number 06437281190
Device Problem High Test Results (2457)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/25/2015
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).
 
Event Description
The customer received a questionable free thyroxine (ft4) result for one patient sample.The result from the cobas 8000 (e602) analyzer serial number (b)(4) was 1.84 ng/dl which was reported outside the laboratory.Based on this result, the physicians decided to stop l-thyroxine treatment and repeat the thyroid levels.The result from the abbott architect was 1.15 ng/dl which was reported.It was unclear if this result was generated from the same patient sample.The patient's current condition was not provided.No adverse event was reported.
 
Manufacturer Narrative
Sample from the patient was submitted for investigation and no specific root cause for the issue with ft4 was found.The customer's results were reproduced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FT4, FREE THYROXINE
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key5213456
MDR Text Key30857862
Report Number1823260-2015-04495
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K961489
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 01/05/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2016
Device Catalogue Number06437281190
Device Lot Number183473
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Event Location Laboratory
Initial Date Manufacturer Received 10/22/2015
Initial Date FDA Received11/10/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/05/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age081 YR
-
-