ROCHE DIAGNOSTICS FT4, FREE THYROXINE; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Catalog Number 06437281190
Device Problem Incorrect Or Inadequate Test Results (2456)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/19/2015
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).
 
Event Description
The customer reported that they received questionable results for one patient sample tested for thyroid parameters.The sample had erroneous results for ft4.The date of the event was asked for, but not provided.The sample was initially tested at the customer site on an e602 analyzer.During investigations, the patient sample was tested on a centaur analyzer, an e170 analyzer, and an e411 analyzer.Refer to the attachment for the specific patient result values.It was asked, but it is unknown which patient results, if any, were reported outside of the laboratory.It was asked, but it is not known if the patient was adversely affected.No adverse events were alleged.The e602 analyzer serial number used at the customer site was asked for, but not provided.The e170 analyzer used for investigation purposes was serial number (b)(4).The ft4 reagent lot number used on this analyzer was 180539, with an expiration date of october 2015.The e411 analyzer used for investigation purposes was serial number (b)(4).The ft4 reagent lot number used on this analyzer was 184927, with an expiration date of march 2016.Based on the information provided, a specific root cause could not be determined.There was not enough sample volume remaining for further investigation.Given the different setups of all assays, the antibodies used, and the variances in reference methods, differences in values may occur when comparing these assays from different vendors.For thyroid parameters, age, gender, and other characteristics are to be considered when analyzing measured values.A general reagent issue can most likely be excluded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FT4, FREE THYROXINE
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key5214267
MDR Text Key30895714
Report Number1823260-2015-04509
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K961489
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Health Professional
Type of Report Initial
Report Date 11/10/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/10/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number06437281190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Event Location Laboratory
Date Manufacturer Received10/23/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-