• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MAPLE GROVE INNOVA¿; STENT, SUPERFICIAL FEMORAL ARTERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MAPLE GROVE INNOVA¿; STENT, SUPERFICIAL FEMORAL ARTERY Back to Search Results
Model Number H74939293052030
Device Problem Bent (1059)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/21/2015
Event Type  Injury  
Manufacturer Narrative
(b)(4).Device is a combination product.(b)(4).Device evaluated by mfr: the complaint device was not returned for analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
 
Event Description
It was reported that stent damage occurred.The very tight target lesion was located in the superficial femoral artery.A 5x200x130 innova stent was advanced to treat the lesion.The stent was delivered deep and distally however, when the device was implanted, it was noted that the stent got stretched proximally.Another stent was advanced to cover the damage stent and the procedure was completed.No patient complications were reported and the patient's status was fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNOVA¿
Type of Device
STENT, SUPERFICIAL FEMORAL ARTERY
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key5220930
MDR Text Key31060601
Report Number2134265-2015-07775
Device Sequence Number1
Product Code NIP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140028
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 10/22/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2018
Device Model NumberH74939293052030
Device Catalogue Number39293-05203
Device Lot Number18095733
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/22/2015
Initial Date FDA Received11/12/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/16/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-