Catalog Number 06437281190 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/20/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event occurred in (b)(6).
|
|
Event Description
|
The customer complained of erroneous results for 1 patient sample tested for free thyroxine (ft4 ii).The erroneous result was reported outside of the laboratory.The initial ft4 ii result from the e602 analyzer was 0.68 ng/dl.The patient also had a tsh result at this time of 6.20 mu/l.Both results were reported outside of the laboratory.An aliquot was made from the primary tube and the sample was repeated on an abbott architect instrument and the ft4 result was 1.16 ng/dl.No adverse event occurred.The e602 analyzer serial number was (b)(4).
|
|
Manufacturer Narrative
|
Quality controls were acceptable.Based on the available data, a general reagent issue can most likely be excluded.A pre-analytical issue cannot be excluded.
|
|
Manufacturer Narrative
|
The sample was submitted for investigation.The ft4 values generated at the customer site could not be confirmed with the measurements received during the investigation.The ft4 values were in the normal reference range.No reagent specific issue could be identified in the investigated sample.Upon visual inspection of the sample, no clots or fibrin filaments were detected.It was recommended to check the pre-analytic procedure at the customer site.
|
|