MAUDE Adverse Event Report: ROCHE DIAGNOSTICS FT4, FREE THYROXINE; RADIOIMMUNOASSAY, FREE THYROXINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS FT4, FREE THYROXINE; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Catalog Number 06437281190
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/20/2015
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).
 
Event Description
The customer complained of erroneous results for 1 patient sample tested for free thyroxine (ft4 ii).The erroneous result was reported outside of the laboratory.The initial ft4 ii result from the e602 analyzer was 0.68 ng/dl.The patient also had a tsh result at this time of 6.20 mu/l.Both results were reported outside of the laboratory.An aliquot was made from the primary tube and the sample was repeated on an abbott architect instrument and the ft4 result was 1.16 ng/dl.No adverse event occurred.The e602 analyzer serial number was (b)(4).
 
Manufacturer Narrative
Quality controls were acceptable.Based on the available data, a general reagent issue can most likely be excluded.A pre-analytical issue cannot be excluded.
 
Manufacturer Narrative
The sample was submitted for investigation.The ft4 values generated at the customer site could not be confirmed with the measurements received during the investigation.The ft4 values were in the normal reference range.No reagent specific issue could be identified in the investigated sample.Upon visual inspection of the sample, no clots or fibrin filaments were detected.It was recommended to check the pre-analytic procedure at the customer site.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FT4, FREE THYROXINE
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key5221080
MDR Text Key31077977
Report Number1823260-2015-04533
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K961489
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 12/29/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number06437281190
Device Lot Number183473
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Event Location Laboratory
Initial Date Manufacturer Received 10/22/2015
Initial Date FDA Received11/12/2015
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received11/17/2015
12/29/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age072 YR
-
-