The freedom ac power supply is a component that enables the freedom driver to be plugged into an external power source.The customer reported that the freedom ac power supply had intermittent functioning loose connection.The customer also reported that the ac power supply was replaced, and there was no reported patient impact.The ac power supply was returned to syncardia for evaluation.Visual inspection revealed a missing connector o-ring, which provides a seal between the ac power supply connector and power adaptor to prevent dust and moisture from interfering with the connection.The missing o-ring did not prevent the ac power supply from properly connecting to a power adaptor.During the investigation, the ac power supply was evaluated and met all functional and electrical test requirements.There was no evidence of a loose connection fit or interruption in performance when it was connected to a freedom power adaptor during the evaluation.The customer-reported issue could not be duplicated or confirmed.The ac power supply performed as intended, and there was no evidence of a malfunction.The ac power supply was taken out of service.The reported issue posed a low risk to the patient because it did not prevent the freedom driver from performing its life-sustaining functions.The freedom driver continued to function as intended.The freedom driver has a redundant power source of onboard batteries, and patients are provided with backup ac power supplies.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.
|