MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, INC. SYNCARDIA FREEDOM AC POWER SUPPLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, INC. SYNCARDIA FREEDOM AC POWER SUPPLY Back to Search Results
Catalog Number 295050-001
Device Problems Loose or Intermittent Connection (1371); Device Stops Intermittently (1599); Device Operates Differently Than Expected (2913)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 11/06/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
The freedom ac power supply is a component that enables the freedom driver to be plugged into an external power source.The customer reported that the freedom ac power supply has intermittent functioning loose connection.This alleged failure mode poses a low risk to the patient because it did not prevent the freedom driver from performing its life-sustaining functions.The freedom driver continued to function as intended.The freedom driver has a redundant power source of onboard batteries.The freedom ac power supply will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.
 
Manufacturer Narrative
(b)(4) fup 1.
 
Event Description
The freedom ac power supply is a component that enables the freedom driver to be plugged into an external power source.The customer reported that the freedom ac power supply had intermittent functioning loose connection.The customer also reported that the ac power supply was replaced, and there was no reported patient impact.The ac power supply was returned to syncardia for evaluation.Visual inspection revealed a missing connector o-ring, which provides a seal between the ac power supply connector and power adaptor to prevent dust and moisture from interfering with the connection.The missing o-ring did not prevent the ac power supply from properly connecting to a power adaptor.During the investigation, the ac power supply was evaluated and met all functional and electrical test requirements.There was no evidence of a loose connection fit or interruption in performance when it was connected to a freedom power adaptor during the evaluation.The customer-reported issue could not be duplicated or confirmed.The ac power supply performed as intended, and there was no evidence of a malfunction.The ac power supply was taken out of service.The reported issue posed a low risk to the patient because it did not prevent the freedom driver from performing its life-sustaining functions.The freedom driver continued to function as intended.The freedom driver has a redundant power source of onboard batteries, and patients are provided with backup ac power supplies.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM AC POWER SUPPLY
Type of Device
AC POWER SUPPLY
Manufacturer (Section D)
SYNCARDIA SYSTEMS, INC.
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
carole marcot
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key5224707
MDR Text Key31463872
Report Number3003761017-2015-00388
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Remedial Action Replace
Type of Report Initial,Followup
Report Date 11/06/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number295050-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/11/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/06/2015
Initial Date FDA Received11/15/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/02/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/24/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age45 YR
-
-