Device 1 of 2: reference mfr.Report#: 1627487-2015-05596.It was reported the patient was experiencing intermittent, uncomfortable overstimulation when stimulation was enabled.An impedance check revealed no anomalies.On (b)(6) 2015, the patient underwent surgical intervention.During the procedure, the patient's lead was found to be damaged at the tail end.Allegedly, the doctor believes a lead contact was damaged and lodged in the header of the ipg.As a result, the patient's lead and ipg were explanted and replaced (with different models).Surgical intervention resolved the patient's issue.
|