(b)(4).Two pictures were received for analysis.During the visual inspection, one of the pictures shows the subassembly p/n (b)(4) assembled with an aquapak sterile water bottle.The second picture shows the insert ((b)(4)) and can be observed the batch number (b)(4).No other issues were found.No functional inspection can be performed since the device sample is not available for evaluation.Failure mode could not be duplicated since it is unknown the reason for the adaptor leak.However, as an additional test, oxygen entrainment testing was performed to 30 subassembly (p/n: (b)(4)) production samples that were taken randomly from adaptors assembly line.These components are part of catalog number 30228 batch (b)(4).Catalog number 30228 uses the same subassembly (b)(4) than catalog number 403128 related to this customer complaint.During the testing and visual inspection no issues or discrepancies were found than can lead to the condition reported by the customer.As an additional test, one of the subassembly p/n (b)(4) production sample was assembled with a sterile water concha mini bottle and tested according oxygen entrainment testing.During the testing no issues or discrepancies were found that can lead to the condition reported by the customer.The device history record review did not show any issues or discrepancies which could potentially relate to this complaint.No conclusion can be established at this time based on the lack of the device sample.The physical sample is necessary in order to perform a proper investigation.If the device sample becomes available this complaint will be reopened.Customer complaint cannot be confirmed.However, the personnel from the assembly line were notified in order to keep the awareness of this type of issue.
|