• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS FT4, FREE THYROXINE; RADIOIMMUNOASSAY, FREE THYROXINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS FT4, FREE THYROXINE; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Catalog Number 06437281160
Device Problem High Test Results (2457)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/02/2015
Event Type  malfunction  
Manufacturer Narrative
Na.
 
Event Description
The field application specialist reported the customer had an ongoing issue with questionable free thyroxine (ft4) results.The samples were processed by the modular preanalytic analyzer (mpa).The specific number of patients involved was unknown.Data was provided for one patient sample.The results on cobas e602 analyzer serial number (b)(4) were 3.4 and 3.4ng/dl.The sample was repeated on a siemens centaur analyzer and the result was 2.3 ng/dl.This result was reported outside the laboratory.There was no adverse event.The customer declined a service visit.
 
Manufacturer Narrative
A potential interference factor was suspected.As no sample material could be provided for investigation, a specific root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FT4, FREE THYROXINE
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key5233685
MDR Text Key31570709
Report Number1823260-2015-04586
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K961489
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 11/19/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/18/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2016
Device Catalogue Number06437281160
Device Lot Number18689401
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Event Location Laboratory
Date Manufacturer Received11/10/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age016 YR
-
-