Model Number M004EPM4500N40
Device Problem Kinked (1339)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/31/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4) device evaluated by manufacturer - the device has a kink at 14mm from the tip while in the neutral position between ring 1 and 2.In addition the ring #1 has broken adhesive and fluids under it.The steering knob and the tension control knob functioned properly on both lock and unlock positions.The curve shape was tested and both curves are not placed in the template shaded area.The device failed the dimensional test.X ray analysis revealed the center support is kinked.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
 
Event Description
Reportable based on analysis completed on 23oct2015.It was reported that the catheter was kinked.During a flutter procedure, the intellatip mifi xp temperature ablation catheter the device was kinked.The catheter was removed and the procedure was completed with a non-bsc device.However; returned device analysis revealed broken adhesive between the electrodes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLATIP MIFI¿ XP TEMPERATURE ABLATION CATHETER
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (HEREDIA)
302 parkway
la aurora
heredia
CS 
Manufacturer (Section G)
BOSTON SCIENTIFIC - COSTA RICA (HEREDIA)
302 parkway
la aurora
heredia
CS  
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key5234104
MDR Text Key31555349
Report Number2134265-2015-07771
Device Sequence Number1
Product Code OAD
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
P020025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 10/23/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/18/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/21/2018
Device Model NumberM004EPM4500N40
Device Catalogue NumberEPM4500N4
Device Lot Number17614450
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/17/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/23/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/05/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-