Catalog Number MA-D20
Device Problems Break (1069); Loose or Intermittent Connection (1371); Temperature Problem (3022)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The actual device was returned for evaluation.Reliability engineering evaluated the device and found that the bearings were worn and loose causing the device to run over the temperature specification.Therefore the reported condition was confirmed.The assignable root cause was determined to be due to worn out bearings.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the attachment device was broken.During in-house engineering it was observed that the bearings were worn and loose causing the device to run over the temperature specification.The event was not reported to have occurred during surgery.It was not reported if there were any delays in a surgical procedure or if a spare device was available.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINIMAL ACCESS ATTACHMENT
Type of Device
DRILLS, BURRS, TREPHINES & ACCESSORIES - ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key5237851
MDR Text Key31932371
Report Number1045834-2015-12563
Device Sequence Number1
Product Code HBE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK042783
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 10/26/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/19/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberMA-D20
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/16/2015
Date Manufacturer Received11/18/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/29/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-