DEPUY SYNTHES POWER TOOLS COMPACT AIR DRIVE II; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESSORIES AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 511.701 |
Device Problems
Mechanical Problem (1384); Physical Resistance (2578); Mechanical Jam (2983)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the motor was seized, jammed and heavy moving.It was also noted that the motor was blocked and the device was leaking.The assignable root cause was determined to be due to normal wear.If additional information should become available, a supplemental medwatch report will be sent accordingly.
|
|
Event Description
|
It was reported from (b)(6) that during service and evaluation, it was observed that the motor seized, jammed and was heavy moving on the compact air drive device.It was noted in the service order that the device was leaking.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|