Siemens was notified on (b)(6) 2015 that the customer recently received an upgrade to syngo rt therapist 4.3.1 mr2 and reported that there is no longer a hard coded pause with a message about moving components if any beam has changes/differences in parameters (example: different table isocenter positions 0 to 270) from one beam to another in an auto-sequence group.But, in mr2 there is no pause or message for the customer to be aware of moving parts; treatment continues without pause.In rare cases of extreme patient positioning there is potential for collision.However, there is no report of injury or mistreatment to a patient.
|
Siemens' hardware investigation shows that the described behavior of the linac control console works as specified.However, the syngo rt therapist 4.3.1_mr2 software update introduced a change regarding the handling of relative/absolute table setup whereby the system no longer generates a warning message about moving components.A customer advisory notice dated january 27, 2016 was distributed to customers, informing them of the workaround to add a manual pause between the treatment fields in order for the system to interrupt/pause the
treatment to move the table and the gantry manually.The reported software behavior will be corrected in a future update, syngo rt therapist 4.3.1_mr3.
|