BIOTRONIK SE & CO. KG LUMAX 340 HF-T; CRT-D Back to Search Results
Model Number 355263
Device Problem Insufficient Information (3190)
Patient Problem Patient Problem/Medical Problem (2688)
Event Date 06/05/2015
Event Type  Injury  
Event Description
This device was replaced because the patient had a traumatic event to the area of where his device was located.The hospital retained the device.Should additional information be received, this file will be updated.
 
Manufacturer Narrative
We received your event description for the above mentioned device and would like to thank you for supporting our post-market surveillance.As of today, the medical device is not available for analysis, therefore the device itself could not be investigated.The information you provided has been entered into our quality system as a complaint.These types of complaints are used to evaluate systems and device performance throughout our organization and help to maintain and improve the performance of our devices.Should additional relevant information or the device itself become available, the investigation will be updated.
 
Manufacturer Narrative
The device was received and analyzed.The memory content of the icd was inspected,showing a normal device function while implanted and in service.In a next step, a sensing test was performed and the device sensed the attached heart signals free of noise.The icd was subjected to an electrical analysis.The ability of the device to deliver therapies was verified.The anti-bradycardia pacing pulses proved to be normal and in amplitude and frequency as programmed.A fibrillation signal was applied and the device delivered a defibrillation shock as specified, documenting a correct sensing and shock delivery.In particular, the specified energy level was reached.In conclusion, the device proved to be fully functional.The analysis of the memory content showed a normal device behavior while the device was implanted and in service.There was no indication of a device malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUMAX 340 HF-T
Type of Device
CRT-D
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin D-123 59
GM  D-12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key5246178
MDR Text Key32014044
Report Number1028232-2015-04143
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P050023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date 11/09/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number355263
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/09/2015
Initial Date FDA Received11/24/2015
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received12/04/2015
07/07/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/18/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age86 YR
-
-