Model Number 3612-100 |
Device Problem
Device Operates Differently Than Expected (2913)
|
Patient Problem
Pneumothorax (2012)
|
Event Date 11/13/2015 |
Event Type
Injury
|
Manufacturer Narrative
|
We are in the process of performing the investigation and will submit the follow-up report once the evaluation is completed.Related mdr 1219977-2015-00333.
|
|
Event Description
|
Report received stated patient had a pneumothorax.A drain was attached to the patient and only a bubble or two would come out of the air leak chamber and then would stop.Under x-ray it was observed that air was building up in the pleural space.The nurse unplugged the drain from the patient and air leak chamber was bubbling rapidly.They had to use a syringe to alleviate air from the pneumothorax being building up.The drain was being used with a 6fr pigtail catheter.
|
|
Manufacturer Narrative
|
Engineering evaluation: the unit was returned and evaluated.The returned drain was set-up and evaluated per the instructions for use (ifu) and standard procedures.The drains suction was set at -20 cmh2o and it measured within acceptable limits (-19.87 cmh2o).Conclusion: there were no issues found during the evaluation of the returned drain.The probable cause was most likely a blocked catheter caused by a clot or the catheter eyelets being against the chest wall.
|
|
Search Alerts/Recalls
|