• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD RESPIRATORY HUMIDIFIER; BTT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD RESPIRATORY HUMIDIFIER; BTT Back to Search Results
Model Number MR850
Device Problem No Audible Alarm (1019)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/28/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Method: the complaint spare pcb assembly was returned to fisher & paykel healthcare in (b)(6) for investigation.It was visually inspected and the coil resistance of the buzzer was measured.Results: visual inspection revealed no signs of impact damage.The coil resistance of the buzzer on the returned spare pcb assembly was noted to be greater than the resistance of a working buzzer.A lot check revealed no other complaints of this nature for lot 150915.Conclusion: we are unable to determine what may have caused the damage observed on the buzzer of the returned spare pcb assembly.All spare pcb assemblies undergo an operator assisted test for the function of the buzzer prior to being released for distribution.This suggests the reported damage occured after the spare pcb assembly was released for distribution.Our mr850 product technical manual contains a maintenance schedule which instructs the user to conduct annual visual checking and performance testing of the mr850 heater base.In addition the product technical manual also states that "all servicing procedures should be followed by a humidifier test, and an electrical safety test to ensure proper operation".The mr850 is equipped with visual alarm indicators in addition to the audible alarm.
 
Event Description
The biomedical engineer from a healthcare facility in (b)(6), reported that an mr850 respiratory humidifier's spare pcb assembly was not making any noise upon installing the pcb into the humidifier.No patient consequence was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRATORY HUMIDIFIER
Type of Device
BTT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
raymond yan
15365 barranca parkway
irvine, CA 92618-2216
9494534000
MDR Report Key5252563
MDR Text Key32201548
Report Number9611451-2015-00503
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K073706
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 10/28/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMR850
Device Catalogue NumberMR850
Device Lot Number150915
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/28/2015
Initial Date FDA Received11/26/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/15/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-