Model Number 2AF284
Device Problem Insufficient Information (3190)
Patient Problem Nerve Damage (1979)
Event Date 11/17/2015
Event Type  Injury  
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.(b)(4).
 
Event Description
It was reported that during a cryo ablation procedure, the patient experienced a phrenic nerve injury after an ablation to the vein.The double-tap technique was performed.The procedure was aborted and the patient was extubated.It was noted that physician planned a sniff test prior to the patient's discharge.Follow up was conducted with no further information regarding the patient's condition.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC FRONT ADVANCE CARDIAC CRYOABLATION CATHETER
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
16771 chemin ste-marie
kirkland H9H 5 H3
CA  H9H 5H3
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key5255514
MDR Text Key32353657
Report Number3002648230-2015-00433
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100010/S015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/17/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/04/2016
Device Model Number2AF284
Device Catalogue Number2AF284
Device Lot Number57492
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/17/2015
Initial Date FDA Received11/30/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received02/01/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/05/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age00059 YR
Patient Weight112
-
-