• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION EON MINI IPG, 16-CHANNEL RECHARGEABLE; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION EON MINI IPG, 16-CHANNEL RECHARGEABLE; SCS IPG Back to Search Results
Model Number 3788
Device Problems Nonstandard Device (1420); No Device Output (1435); Charging Problem (2892); Communication or Transmission Problem (2896)
Patient Problem Inadequate Pain Relief (2388)
Event Type  Injury  
Manufacturer Narrative
Field advisories: 1627487-07262012-002-r; 1627487-12192011-003-r.This ipg serial number was included in field advisories.(b)(4).Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.(b)(4).
 
Event Description
It was reported the patient was unable to establish communication between his ipg and charger.A new charger was forwarded to the patient, which did not resolve the issue.It was also reported the patient's programmer reflected a communication error message.In addition, the patient stated his recharging time gradually increased.It was noted the patient has multiple unrelated health issues and uses a morphine pump.The patient was without stimulation.The sjm representative met with the patient and confirmed the issues.Subsequently, the patient underwent surgical intervention on (b)(6) 2015 where the ipg was removed and replaced with a different model.The patient reported effective stimulation therapy postoperative and the reported issue is now resolved.The event date is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON MINI IPG, 16-CHANNEL RECHARGEABLE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston road
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston road
plano TX 75024
MDR Report Key5259175
MDR Text Key32451374
Report Number1627487-2015-23704
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
Report Date 11/06/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2010
Device Model Number3788
Device Lot Number175163
Was Device Available for Evaluation? No
Date Manufacturer Received11/06/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/12/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Removal/Correction Number1627487-05242011-002-R
Patient Sequence Number1
Patient Outcome(s) Other;
-
-