The customer questioned low results with pediatric samples tested for gluco-quant glucose/hk (glu) with reagent lots r1 reagent - (b)(4) and r2 reagent - (b)(4) on a modular p analyzer.These reagents have been in use since (b)(6) 2015.Data was provided for (b)(6) pediatric samples.Of the data provided, (b)(6) patient samples were erroneous.The samples were repeated with new reagent lots of r1 reagent - (b)(4) and r2 reagent - (b)(4) on a 2nd modular p analyzer.The customer thinks the results from the new reagent lots are correct.The initial low results did not fit the clinical picture for these patients.The erroneous results were reported outside of the laboratory.The corrected results were provided to the physician.It is not clear whether the repeat results were performed on the same day as the initial results or a week later.See the attachment to the medwatch for the patient results.No adverse event occurred.The modular p analyzer serial number was (b)(4).It was noted that the customer has had no issues since changing to the new reagent lots on (b)(6) 2015.It was also noted that the customer had no issue with adult samples tested for glu.The issue only occurred with pediatric samples.
|