Catalog Number 06437281190
Device Problem High Test Results (2457)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/20/2015
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).
 
Event Description
The customer received questionable free thyroxine (ft4) results for one patient sample from a cobas e602 analyzer.The specific date of testing was not provided.The sample was submitted for investigation and was tested on a centaur analyzer, an analytical e module analyzer (e170), and a cobas e411 analyzer.Refer to the attachment to the medwatch for all patient data.Information concerning if any result was reported outside of the laboratory was requested, but was not provided.No adverse event was reported.As insufficient sample material remained for further testing, a specific root cause could not be determined.From the provided data, a general reagent issue was not suspected.Upon comparing results generated by different types of analyzers, variances can be expected due to the different setups of all assays, the antibodies used, and variances in the reference methods.For thyroid parameters, age, gender, and other characteristics should be considered when analyzing results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FT4, FREE THYROXINE
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key5261193
MDR Text Key32511101
Report Number1823260-2015-04663
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K961489
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Health Professional
Type of Report Initial
Report Date 12/02/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/02/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number06437281190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Event Location Laboratory
Date Manufacturer Received11/20/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-