Model Number PARADYM RF SONR CRT-D 9770 |
Device Problem
Over-Sensing (1438)
|
Patient Problem
Shock (2072)
|
Event Date 11/12/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device model involved in this mdr report is not approved in the united states; however, it is similar to a device manufactured by sorin that was cleared or approved by fda for marketing in the united states.
|
|
Event Description
|
Reportedly, the inappropriate shocks were delivered due to oversensing on rv lead.
|
|
Event Description
|
Reportedly, the inappropriate shocks were delivered due to oversensing on rv lead.
|
|
Manufacturer Narrative
|
Preliminary analysis of the available data did not reveal any programmer software issue.The reported behavior is most probably related to t-wave oversensing.
|
|
Event Description
|
Reportedly, the inappropriate shocks were delivered due to oversensing on rv lead.
|
|