MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. ATTUNE SHIM SZ9 14MM; KNEE INSTRUMENT/TRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. ATTUNE SHIM SZ9 14MM; KNEE INSTRUMENT/TRIAL Back to Search Results
Catalog Number 254500696
Device Problems Break (1069); Component Missing (2306)
Patient Problem No Information (3190)
Event Date 11/18/2015
Event Type  malfunction  
Manufacturer Narrative
This complaint is under investigation.Depuy will notify the fda of the results of the investigation once it has been completed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Event Description
A large piece of the shim has broken off.This case has been reported by the loan kit technician during loan kit inspection.
 
Manufacturer Narrative
This complaint remains under investigation.Depuy will notify the fda of the results of the investigation once it has been completed.
 
Manufacturer Narrative
Examination of the returned device confirms the reported event of trial damage/breakage.The observed cracking/breaking is initiating around the metal insert of the trial.A pra 103121982 (preliminary risk assessment) meeting was held to discuss a product escalation out of australia concerning the attune shims.The team has determined that there is no additional patient risk associated with this issue.Although cracks are occurring, this failure mode does not lead to a patient harm of infection.The rate of infection for attune is within the state rate in the risk management report and is statistically similar to the rate of infection for the total knee class.Although a definitive root cause is not known, a combination user technique, improper technique or misuse, and the use of contraindicated chemicals (non-compliant to cleaning guidelines), may lead to weakening of the material, as seen in the returned device.Based on the performed investigation and pra 103121982 determination of no additional patient risk, corrective action is not indicated.Continue to monitor via sep-419.Depuy considers the investigation closed at this time.Should the additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE SHIM SZ9 14MM
Type of Device
KNEE INSTRUMENT/TRIAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46581
5743725905
MDR Report Key5263321
MDR Text Key32838645
Report Number1818910-2015-36444
Device Sequence Number1
Product Code HWT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup,Followup
Report Date 12/21/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/03/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254500696
Device Lot NumberBFA0N98
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received12/23/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/27/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-