Additional/updated information was added to reflect the device receiving an expanded evaluation.However, subsequent testing could not verify the complaint.The foot section and all motors were visually inspected to be in good condition.The cables and cable connectors were visually inspected and were found to be free from cuts, breaks, bends or other damage.The power supply and hand pendant were not returned, so a set of test components was used to assess the functionality of the foot section.All motors functioned correctly, and the foot section was able to raise and lower as intended.No sparks were observed, and no shock was detected when the frame, motors, and cables were physically touched.A multimeter was used to measure the resistance of each motor, which all displayed an open circuit, meaning that no short circuits were found.The underlying cause of the complaint issue could not be determined.
|