ZOLL MEDICAL CORPORATION M SERIES; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number MSERIES BI-PHASIC
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
 
Event Description
Complainant alleged that during a routine shift check by a clinician, the device displayed a unknown "pacer fault" message.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Manufacturer Narrative
The malfunction was observed during incoming testing; however after disassembling the device to determine root cause, the malfunction was no longer seen.The customer declined repair and the device was returned not certified and labeled not for clinical use.No trend is associated with reports of this type.
 
Event Description
Complainant alleged that during a routine shift check by a clinician, the device displayed "pacer fault 116" and "pacer disabled" messages.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M SERIES
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key5268615
MDR Text Key32933592
Report Number1220908-2015-03180
Device Sequence Number1
Product Code DRO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K990762
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/20/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/04/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberMSERIES BI-PHASIC
Device Catalogue NumberM SERIES
Other Device ID Number00847946003937
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received11/20/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2001
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-