BOSTON SCIENTIFIC - MAPLE GROVE INNOVA¿; STENT, SUPERFICIAL FEMORAL ARTERY Back to Search Results
Model Number H74939293071030
Device Problem Migration or Expulsion of Device (1395)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/12/2015
Event Type  Injury  
Manufacturer Narrative
Age at time of event: 18 years or older.(b)(4).Device is a combination product.(b)(4).Device evaluated by mfr.: the device was not returned for evaluation.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical/ procedural factors.(b)(4).
 
Event Description
It was reported that stent migration occurred.The 90% stenosed, 7mm target lesions were located in the right common and right external iliac arteries.A 7 x 100 x 130 innova stent was selected for use, advanced and deployed to treat the lesion.However, post stent deployment, the stent jumped into the aorta.The physician then deployed the balloon proximal to the stent and dragged the innova⠓tent downward to the location where the stent was intended to be implanted.A balloon expandable stent was advanced to the proximal portion of the migrated innova stent, secured the innova stent to the vessel wall and completed the procedure.No patient complications were reported and the patient's status was fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNOVA¿
Type of Device
STENT, SUPERFICIAL FEMORAL ARTERY
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key5274445
MDR Text Key32933865
Report Number2134265-2015-08299
Device Sequence Number1
Product Code NIP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140028
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 11/12/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/08/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2018
Device Model NumberH74939293071030
Device Catalogue Number39293-07103
Device Lot Number18137886
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/12/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/29/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-