• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET ORTHOPEDICS COMP RVS HMRL TI TRAY 44MM; PROSTHESIS, SHOULDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET ORTHOPEDICS COMP RVS HMRL TI TRAY 44MM; PROSTHESIS, SHOULDER Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem Bone Fracture(s) (1870)
Event Date 11/18/2015
Event Type  Injury  
Manufacturer Narrative
Current information is insufficient to permit a conclusion as to the cause of the event.Review of device history records show that lot released with no recorded anomaly.There are warnings in the package insert that state that this type of event can occur: under warnings, it states, ¿improper selection, placement, positioning, alignment and fixation of the implant components may result in unusual stress conditions which may lead to subsequent reduction in the service life of the prosthetic components.¿ under possible adverse effects, number 10 states, "fatigue fracture of component can occur as a result of loss of fixation, strenuous activity, malalignment, trauma, non-union, or excessive weight.".
 
Event Description
It was reported that patient underwent a reverse shoulder arthroplasty on (b)(6) 2011.Subsequently, patient was revised on (b)(6) 2015 due to humeral tray fracture.No further information has been provided.
 
Manufacturer Narrative
This follow-up report is being submitted to relay corrected and additional information.Device was not returned for analysis; therefore, a product evaluation could not be conducted.Dhr was reviewed and no discrepancies were found.Review of the complaint history determined that no further action is required as no were trends identified.Root cause was undetermined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMP RVS HMRL TI TRAY 44MM
Type of Device
PROSTHESIS, SHOULDER
Manufacturer (Section D)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key5275208
MDR Text Key32981569
Report Number0001825034-2015-04922
Device Sequence Number1
Product Code PAO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK080642
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/08/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/31/2021
Device Model NumberN/A
Device Catalogue Number115340
Device Lot Number579670
Other Device ID NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/23/2017
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/16/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-