(b)(4) method: the complaint mr850 respiratory humidifier was not returned to fisher & paykel healthcare for evaluation.Therefore, our investigation is based on the information provided by the hospital, previous similar investigations and our knowledge of the product.Results: it was reported that a third party breathing circuit manufactured by intersurgical was used in the setup at the time of the reported event.The hospital further reported that there was no patient consequence.A lot check could not be carried out as the lot information was not provided.Conclusion: based on the information provided, the condensation issue reported by the customer is most likely related to the use of a third party breathing circuit with the mr850 respiratory humidifier.Condensate in the humidification system is an expected side effect of heated pass-over humidification systems in many conditions, and may vary between light misting to water droplets that form on the wall of cool breathing circuit tubing.The amount of condensate in the ventilation system is influenced by a number of multiple set up and environmental factors including third party breathing circuits not approved by fisher & paykel healthcare.The user instructions that accompany the mr850 respiratory humidifier state: "the use of breathing circuits, chambers or other accessories which are not approved by fisher & paykel healthcare may impair performance or compromise safety." since the reported incident a fisher & paykel healthcare representative has visited the hospital to answer any questions and address any condensation issues.No condensation issues were found during the rep's visit.
|