(b)(4).Device evaluated by mfr.: the device was returned for analysis.The device was returned with the stent partially deployed by 2mm.A protruding wire was observed on the partially deployed unit on the distal side of the stent.There was damage identified to the stent.During the product analysis, the device was flush without issue.A visual and tactile examination found no kinks or other damage along the length of the catheter.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The root cause is not confirmed as there was no evidence of the alleged issue or any anomalies which could have contributed to the reported difficulty.(b)(4).
|
Reportable based on device analysis completed on 27-nov-2015.It was reported that difficulty flushing occurred.A 8x29, 5f, 135cm carotid wallstent was selected for use to treat the lesion.However, during preparation, the physician could not flush the lumen.The procedure was completed with another of the same device.No patient complications were reported.However, returned device analysis revealed partial stent deployment and stent damage.
|