Model Number ADDR01
Device Problem Incorrect Measurement (1383)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/16/2015
Event Type  malfunction  
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.(b)(4).
 
Event Description
It was reported that the right atrial (ra) lead exhibited undersensing.Also, the implantable pulse generator (ipg) appeared to be under reporting the atrial burden.The atrial burden and cardiac compass don't match the histogram data possibly due to post-modeswitch overdrive pacing (pmop) and collection delay.Reprogramming was discussed and the lead and device remain in use.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADAPTA DR
Type of Device
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
ceiba norte industrial park, r
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key5283726
MDR Text Key33482339
Report Number3004209178-2015-24722
Device Sequence Number1
Product Code NVZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P980035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 09/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/28/2014
Device Model NumberADDR01
Device Catalogue NumberADDR01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/16/2015
Initial Date FDA Received12/09/2015
Date Device Manufactured10/04/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age00080 YR
-
-