• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. ATTUNE SHIM SZ9 5MM; KNEE INSTRUMENT/TRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. ATTUNE SHIM SZ9 5MM; KNEE INSTRUMENT/TRIAL Back to Search Results
Catalog Number 254500691
Device Problems Disassembly (1168); Component Missing (2306)
Patient Problem No Information (3190)
Event Date 11/19/2015
Event Type  malfunction  
Manufacturer Narrative
This complaint is under investigation.Depuy will notify the fda of the results of the investigation once it has been completed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Event Description
Shim has returned without steel insert.This case has been reported by the loan kit technician during loan kit inspection.
 
Manufacturer Narrative
The device associated with this report was not returned.Follow up communication for product return was unsuccessful.A complaint database search on the provided product code family showed similar reports for damage/cracking.Previous reports found user technique, misuse, or the use of a contraindicated cleaning chemical, as suspected root causes.The investigation could not verify or identify any product contribution to the reported event with the information provided as the reported device was not returned for evaluation.Based on the inability to identify root cause, the need for corrective action was not indicated.Continue to monitor via (b)(4).Depuy considers the investigation closed at this time.Should the additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Manufacturer Narrative
This investigation has been reopened because the product has been received.Depuy will notify the fda of the results of the investigation once it has been completed.
 
Manufacturer Narrative
Examination of the returned device confirms the reported event of a missing metal insert.A complaint database search on the provided product code family showed similar reports for damage/cracking.Previous reports found user technique, misuse, or the use of a contraindicated cleaning chemical, as suspected root causes.Although a definitive root cause is not known, a combination user technique, improper technique or misuse, and the use of contraindicated chemicals (non-compliant to cleaning guidelines), may lead to weakening of the material, as seen in the returned device.Based on the inability to determine a specific root cause, a need for corrective action is not indicated.Continue to monitor via (b)(4).Depuy considers the investigation closed at this time.Should the additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE SHIM SZ9 5MM
Type of Device
KNEE INSTRUMENT/TRIAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46581
5743725905
MDR Report Key5284826
MDR Text Key33101497
Report Number1818910-2015-37119
Device Sequence Number1
Product Code HWT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup,Followup,Followup
Report Date 01/05/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254500691
Device Lot NumberBFA0MQG
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/05/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/30/2015
Initial Date FDA Received12/10/2015
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Not provided
Supplement Dates FDA Received12/21/2015
01/11/2016
01/13/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/14/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-