Catalog Number S-55-060-120-P6 |
Device Problems
Positioning Failure (1158); Difficult to Remove (1528); Material Deformation (2976)
|
Patient Problem
Foreign Body In Patient (2687)
|
Event Date 11/17/2015 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The stent remains in the vessel.The stent delivery system was reported to be discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
|
Event Description
|
It was reported that during a procedure of the non-tortuous, non-calcified popliteal artery, the supera stent was partially deployed in the vessel, but became caught in an unspecified way and was stretched to the common femoral artery where it became fully deployed but not in the target lesion.The stent remains in a fixed position and is fully apposed to the vessel wall.A second 5.5 x 40 mm supera stent was used to treat the lesion without reported adverse patient effect.Although the procedure was delayed about 15 minutes it was not clinically significant.No additional information was provided.
|
|
Manufacturer Narrative
|
(b)(4).The device is not returning for evaluation as it remains in the patient; therefore, a failure analysis of the complaint device could not be completed.Based on the reported information, there is no indication of a product quality issue with respect to manufacture, design or labeling.A definitive cause for the reported difficulties could not be determined.The foreign body in patient is due to operational context.A review of job traveler revealed no non-conformances that would have contributed to the reported event.The results of the historical data review and query of the electronic complaint handling database revealed no other incidents reported from this lot.Based on the information reviewed, there is no indication the issue was caused by or related to the design, manufacture or labeling of the device.
|
|
Search Alerts/Recalls
|