• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO BCT TRIMA ACCEL; TRIMA ACCEL PLATELET, PLASMA, SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO BCT TRIMA ACCEL; TRIMA ACCEL PLATELET, PLASMA, SET Back to Search Results
Catalog Number 80337
Device Problems Air Leak (1008); Improper or Incorrect Procedure or Method (2017); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/16/2015
Event Type  malfunction  
Manufacturer Narrative
Investigation is in process.A follow-up report will be provided.
 
Event Description
The customer reported that during set up for a donation, they received a 'pressure test error alarm.They found that the pinch clamp on the diversion bag line tubing was not properly closed, and there was air in the diversion bag.The customer was able to remove the air from the diversion bag and continue the procedure.Patient (donor) information and outcome is not available at this time.The disposable kit is not available for return, because it was discarded by the customer.
 
Manufacturer Narrative
A review of the device history record (dhr) for this unit showed no irregularities during manufacturing that were relevant to this issue.The run data file (rdf) was analyzed for this event.Root cause: signals in the rdf showed ¿pressure test error¿ alerts in this procedure occurred when the system was verifying that the sample bag clamp was closed.If the system does not pass this test twice, then the operator is prompted to end the procedure because the system is unable to achieve the required pressure.Based on the customer's report, the sample bag clamp was not fully closed and was not fully occluding the line during the tubing set test,therefore generating the two ¿pressure test error¿ alerts.Correction: a voluntary medical device product recall was released by terumo bct on september 15, 2014 to communicate more information about the "air in the sample bag safety alert".Corrective and preventive action: an internal capa has been initiated to address air in the sample bag incidents.
 
Manufacturer Narrative
This report is being filed to provide additional information.Terumo bct's sales representative followed-up with customer and explained the possible failure modes and the necessary actions to be performed by the customer if air is in the bag, and risks to the donor.
 
Event Description
No patient (donor) was connected at the time of the event.No patient (donor) information is reasonably known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Type of Device
TRIMA ACCEL PLATELET, PLASMA, SET
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
Manufacturer Contact
matthew bickford
10810 w collins ave
lakewood, CO 80215
3032052494
MDR Report Key5290577
MDR Text Key33895085
Report Number1722028-2015-00650
Device Sequence Number1
Product Code GKT
Combination Product (y/n)N
PMA/PMN Number
BK150269
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 12/11/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/01/2017
Device Catalogue Number80337
Device Lot Number09Y0105
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/19/2015
Initial Date FDA Received12/11/2015
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received01/05/2016
03/02/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/04/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-