Model Number OPT318
Device Problem Device Handling Problem (3265)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Method: the complaint opt318 cannula was not made available for our inspection.Our investigation is based on a photograph provided by the hospital and our knowledge of the product.Results: visual inspection of the photograph revealed that the flexitube (cannula tubing) was damaged where it connects to the breathing circuit.The tubing was stretched and broken but the metal spring was still attached to the swivel grip.Conclusion: the damage to the flexitube is likely to have been caused by excessive pulling force being exerted on the tubing.All optiflow junior cannulae are 100% leak and occlusion tested after final assembly and any cannula that fails is discarded.In addition, samples are taken hourly from each run and pull tested to check glue joint strength at the cannula/tube joint, as well as the swivel grip joint.If there are any failures the entire batch is put aside for further investigation.The user instructions illustrate in pictorial format the correct set-up and proper use of the optiflow junior nasal cannula.They also state the following: ensure that all connections are secure during use.Check cannula is undamaged and that the flow path is maintained.Appropriate monitoring must be used at all times.Do not stretch or crush tube.
 
Event Description
A hospital in (b)(6) reported that an opt318 optiflow junior nasal cannula broke at the distal end of the tubing; flow was interrupted and the patient desaturated "for a few seconds" before staff replaced the cannula.The hospital confirmed that there was no patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLOW JUNIOR INTERFACE
Type of Device
CAT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
raymond yan
15365 barranca parkway
irvine, CA 92618-2216
9494534000
MDR Report Key5295836
MDR Text Key33469384
Report Number9611451-2015-00532
Device Sequence Number1
Product Code CAT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Physician
Type of Report Initial
Report Date 11/17/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOPT318
Device Catalogue NumberOPT318
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/17/2015
Initial Date FDA Received12/14/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-