BECKMAN COULTER UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM; COUNTER, DIFFERENTIAL CELL Back to Search Results
Catalog Number 629029
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/22/2015
Event Type  malfunction  
Manufacturer Narrative
The field service engineer (fse) evaluated the instrument on 11/23/2015.The fse found disconnected differential, diff, air mix and temperature control, amtc, sample line from the mix chamber.He replaced the diff amtc sample line and the instrument ran without any leaks and error messages.The repairs were verified per established service procedures.Beckman coulter internal identifier for this report is (b)(4).
 
Event Description
While the field service engineer (fse) was troubleshooting the instrument for the partial clog errors reported by the customer , the fse observed a fluid leak from a unicel dxh 800 coulter cellular analysis system.The volume of the leak was not specified.The leak was contained within the instrument.The fse was wearing personal protective equipment consisting of a laboratory coat and gloves at the time of the event and there was no report of injury or direct exposure to the leak.Erroneous patient results were not generated and there was no change or affect to patient treatment in connection with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s kraemer blvd
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th avenue
kendall FL 33196
Manufacturer Contact
miranda holland
11800 sw 147th avenue
32-s08
kendall, FL 33196
3053802031
MDR Report Key5296777
MDR Text Key34074560
Report Number1061932-2015-01775
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140911
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/22/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/14/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number629029
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/22/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-