MEDTRONIC SOFAMOR DANEK CD HORIZON SPINAL SYSTEM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Catalog Number UNKNOWN
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem No Code Available (3191)
Event Date 11/13/2015
Event Type  Injury  
Manufacturer Narrative
(b)(6).(b)(4).Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.
 
Event Description
It was reported that the patient underwent transforaminal lumbar interbody fusion (tlif) at l4/5 both side by intermuscular approach for lumber degenerative spondylolisthesis.Implant 4.75 was used.Post-op, it was confirmed by x-ray that l4 left placed screw was slipped to lateral.As reported by the local distributer, the doctor considered that he would perform a revision surgery.No patient complications were reported in the report.It was reported that l4 screws on both sides dislodged.In revision surgery, both screws were re-inserted in a different direction and the surgery was completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK
1800 pyramid place
memphis TN 38132
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK
1800 pyramid place
memphis TN 38132
Manufacturer Contact
greg anglin
1800 pyramid place
memphis, TN 38132
9013963133
MDR Report Key5296847
MDR Text Key33470152
Report Number1030489-2015-03430
Device Sequence Number1
Product Code NKB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
UNKNOWN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial
Report Date 12/03/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/03/2015
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age00048 YR
-
-