ST. JUDE MEDICAL, COSTA RICA LTDA SJM TRIFECTA VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number TF-23A
Device Problem Torn Material (3024)
Patient Problems Aortic Regurgitation (1716); Heart Failure (2206)
Event Date 11/16/2015
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
On (b)(6) 2012, this 23 mm sjm trifecta valve was implanted.On (b)(6) 2015, the valve was explanted for severe aortic regurgitation with acute heart failure.At explant a tear was observed at the base of the right and left cusps.A 23 mm tissue valve from another manufacturer was implanted.The patient was reported to be stable postoperatively.
 
Manufacturer Narrative
(b)(4).The results of this investigation concluded a tear was observed in cusp 2, and three tears were observed in cusp 3.Focal inflow pannus was observed on cusp 3.There was no evidence of acute inflammation or calcifications and gram stains were negative for organisms.No evidence was found to suggest the cause of the cuspal tears and pannus formation was due to an intrinsic defect in the valve, as supported by review of the valve's device history record and the analysis performed.The cause of the reported event remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM TRIFECTA VALVE
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela 1897- 4050
CS   1897-4050
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key5299126
MDR Text Key33557322
Report Number3008452825-2015-00120
Device Sequence Number1
Product Code LWR
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P100029
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date12/24/2013
Device Model NumberTF-23A
Device Catalogue NumberTF-23A
Device Lot Number3603229
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/29/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/16/2015
Initial Date FDA Received12/15/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/22/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/26/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age77 YR
-
-