MAUDE Adverse Event Report: BECKMAN COULTER UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECKMAN COULTER UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM Back to Search Results
Catalog Number 629029
Device Problem Leak/Splash (1354)
Patient Problem Not Applicable (3189)
Event Date 11/20/2015
Event Type  malfunction  
Manufacturer Narrative
A field service engineer (fse) evaluated the instrument on (b)(4) 2015.The fse observed a cleaner leak after instrument was shut down.The fse confirmed the diff waste tubing (vl211) was loose causing the leak.The fse replaced the tubing resolving the issue.The repairs were verified per established service procedures.(b)(4).
 
Event Description
The customer reported less than 50 mls of uncontained cleaner fluid leaked from the unicel dxh 800 coulter cellular analysis system.Onto the floor following daily checks.Erroneous patient results were not generated and there was no change or effect to patient treatment in connection to the event.There was no impact to patient results or controls.The customer was wearing personal protective equipment (ppe) consisting of laboratory coat, goggles and gloves at the time of the event and there was no report of injury or biohazard exposure to open wounds or mucous membranes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM
Type of Device
UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM
Manufacturer (Section D)
BECKMAN COULTER
250 s. kraemer blvd
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER, INC.
11800 sw 147th avenue
miami FL 33196 2031
Manufacturer Contact
miranda holland
11800 sw 147th avenue, 32-s08
miami, FL 33196-2031
3053802031
MDR Report Key5300253
MDR Text Key33589949
Report Number1061932-2015-01838
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140911
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/20/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other Health Care Professional
Device Catalogue Number629029
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/20/2015
Initial Date FDA Received12/15/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-