ANGIODYNAMICS PMTA ACCU2I LONG APPLICATOR; MICROWAVE APPLICATOR Back to Search Results
Catalog Number 900-601
Device Problems Bent (1059); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/19/2015
Event Type  malfunction  
Manufacturer Narrative
It was reported that the disposable device was discarded by the user and is not available to be returned to the manufacturer for evaluation.An investigation into the root cause of this incident is currently in progress.The results of the device evaluation will be sent via a follow up medwatch.A review of the device history records was performed for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.(b)(4).
 
Event Description
As reported (b)(6) 2015, patient of unknown age and gender presented for an microwave procedure of the liver.During the procedure, after withdrawal, the treating physician noted the tip of the applicator probe was been.The tip of the applicator had not detached from the shaft of the probe and no wires were exposed.The device was set aside and a new of the same device was used to successfully complete the procedure.It was reported the patient suffered no harm or injury due to the event.The reported disposable device is not available for return to the manufacturer for evaluation as it was disposed of by the user.
 
Manufacturer Narrative
Returned for evaluation was one pmta accu2i intermediate applicator.Visual examination noted that tip of the applicator was bent.The tip did not fully detach.The reported complaint description of the tip of the applicator bending is confirmed.A review of the device history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.Although the exact root cause of the event is unable to be determined, it does not appear to be the result of a manufacturing failure.A possible contributing factor could have been handling damage; i.E.Lateral forces on the applicator tip.The instructions for use, which is supplied to with catalog number, contains a statement "always advance the applicator into the target tissue using axial forces only.Avoid placing lateral forces on the applicator tip during placement or removal" and lateral forces on the applicator tip should be avoided both during insertion and removal.Failure to do so could result in damage to the microwave array, failure of the applicator and injury to the patient." a review of the angiodynamics complaint system noted no trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Complaint # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PMTA ACCU2I LONG APPLICATOR
Type of Device
MICROWAVE APPLICATOR
Manufacturer (Section D)
ANGIODYNAMICS
queensbury NY
Manufacturer (Section G)
ANGIODYNAMICS, INC.
603 queensbury avenue
queensbury NY 12804
Manufacturer Contact
lawrence ryan
03 queensbury avenue
queensbury, NY 12804
5187981215
MDR Report Key5305626
MDR Text Key33838713
Report Number1319211-2015-00377
Device Sequence Number1
Product Code NEY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122762
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/20/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2016
Device Catalogue Number900-601
Device Lot Number15270401
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/23/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/20/2015
Initial Date FDA Received12/09/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received07/11/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-