Model Number V200
Device Problems Thermal Decomposition of Device (1071); Device Emits Odor (1425)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Patient information was requested and was not provided.
 
Event Description
The customer reported a burning smell coming from the rear of the ventilator due to a burnt component.The customer reported the patient was removed from the ventilator and placed on a different ventilator.The customer reported patient involvement with no harm to the patient.
 
Manufacturer Narrative
Date of event: (b)(6) 2015.Date rcvd by mfr: 08/06/2016.Conclusion / root cause: the blower motor controller pcba was tested and no failures were identified.No fault was found on this returned blower motor controller pcba.This report is being submitted as part of the remediation for capa (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V200 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
richa shah
3000 minuteman road
andover, MA 01810
9786871501
MDR Report Key5306684
MDR Text Key34437261
Report Number2031642-2015-02694
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/03/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV200
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/03/2015
Initial Date FDA Received12/17/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/25/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/14/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-