This is being filed to report that during returned device analysis of the steerable guiding catheter, the hemostasis valve bond was found broken, which has the potential to compromise fluid path integrity and lead to air embolism in the patient.It was reported that this was a mitraclip procedure to treat functional mitral regurgitation (mr) with a grade of 4.Clip delivery system (cds) 50804u220 met resistance during advancement into the steerable guide catheter (sgc) and was removed without reported issue.Cds 50817u150 also met resistance during advancement into the same sgc.During withdrawal of the cds, resistance was met with the clip introducer and the clip introducer became damaged.The sgc was replaced with a new sgc.Cds 50804u220 was reinserted and the clip was deployed without reported issue.During preparation of cds 50810u120, the delivery catheter (dc) shaft was noted to be bending when the lock lever was retracted; thus, the cds was not used.A new cds was advanced and the clip was deployed without reported issue.The procedure was completed successfully with mr reduced to 1-2, no reported adverse patient effects, and no reported clinically significant delay in the procedure.There was no additional information provided.The sgc was returned with the hemostasis valve bond broken.
|