DEPUY SYNTHES POWER TOOLS COLIBRI; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 532.001 |
Device Problems
Mechanical Problem (1384); Physical Resistance (2578); Mechanical Jam (2983)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and observed that the motor seized and was rough running.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to normal wear from use and servicing over time.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported from (b)(6) that during service and evaluation, it was observed that the small battery device motor seized and was rough running.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|