Manufacturer Narrative
|
Additional narrative: (b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device motor was seized, rough running and defective.It was also noted that the motor stopped under load.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing.If additional information should become available, a supplemental medwatch report will be sent accordingly.
|
|
Event Description
|
It was reported by (b)(6) that during service and evaluation, it was observed that the motor on the electric pen drive device was seized, rough running and defective.It was also noted that the motor stopped under load.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|