• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECKMAN COULTER UNICEL DXH 800 CELLULAR ANALYSIS SYSTEM; COUNTER, DIFFERENTIAL CELL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECKMAN COULTER UNICEL DXH 800 CELLULAR ANALYSIS SYSTEM; COUNTER, DIFFERENTIAL CELL Back to Search Results
Catalog Number 629029
Device Problem Fluid/Blood Leak (1250)
Patient Problem Not Applicable (3189)
Event Date 11/25/2015
Event Type  malfunction  
Manufacturer Narrative
A field service engineer (fse) evaluated the instrument on 11/25/2015.The fse found that the probe wash collar was broken.The fse replaced the probe wash collar, which repaired the leak.The repairs were verified by the fse.(b)(4).
 
Event Description
The customer reported a bloody leak from the unicel dxh 800 cellular analysis system.The volume of he leak was about 4 cc and was not contained within the instrument.The customer was wearing personal protective equipment (ppe) consisting of gloves, goggles and a laboratory coat at the time of the occurrence and there was no report of injury or direct exposure to the leak.Erroneous patient results were not generated and there was no change or affect to patient treatment in connection with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNICEL DXH 800 CELLULAR ANALYSIS SYSTEM
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 south kraemer blvd.
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th ave
miami FL 33196
Manufacturer Contact
miranda holland
11800 sw 147th ave
32-s08
miami, FL 33196
3053802031
MDR Report Key5316517
MDR Text Key34078386
Report Number1061932-2015-01864
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140911
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Medical Technologist
Type of Report Initial
Report Date 11/25/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number629029
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/25/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-