It was reported that rv oversensing was observed on an egm post-lvad implant.It was noted that adjusting the lvad frequency resulted in an increase in the amplitude of noise.The atrial and rv lead impedances were 310 and 550 ohms; however, x-ray showed the atrial lead had dislodged.Tachy therapy was disabled and programming changes were made.With the new programming changes, rv oversensing was no longer observed; however, crosstalk was present.Additional programming changes were made, which resolved the crosstalk oversensing.The patient will continue to be monitored.
|