No conclusion code available.Final analysis found the reported inability to communicate with the device was confirmed in the laboratory and was due to premature battery depletion.Based on all the available parameter and usage information, device longevity was found to below the expected limits.With an external power supply attached, the device functioned normally.No high current was detected during testing and the power consumption was normal.The original battery was returned to the vendor for further evaluation and analysis could not conclusively determine a root cause for the premature battery depletion.
|
Premature battery depletion was confirmed by analysis.Bench testing on the device was performed, and no sources of high current were noted.In further analysis of the battery, lithium clusters were observed, but at the time of analysis the clusters did not appear to be in a location or size that would be sufficient to cause an internal short of the battery.As a result, the cause of the premature battery depletion could not be conclusively determined.However, from these analyses, in the absence of other root causes, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.
|