MAUDE Adverse Event Report: DATEX-OHMEDA, INC. AESTIVA 7100; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATEX-OHMEDA, INC. AESTIVA 7100; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Device Inoperable (1663)
Patient Problem No Patient Involvement (2645)
Event Date 11/24/2015
Event Type  malfunction  
Manufacturer Narrative
No report of patient involvement.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.A ge healthcare service representative performed a checkout of the equipment and confirmed the reported complaint.The control board was replaced, and the unit was returned to service.
 
Event Description
The hospital reported the unit alarmed 'vext_ref out of range'.There was no report of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AESTIVA 7100
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda drive
madison WI 53718
Manufacturer (Section G)
DATEX-OHMEDA, INC.
3030 ohmeda drive
madison WI 53718
Manufacturer Contact
joseph seliga
3000 n grandview blvd.
waukesha, WI 53188
MDR Report Key5318225
MDR Text Key34119308
Report Number2112667-2015-00445
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K000706
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial
Report Date 11/24/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/24/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/05/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-