MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. (CRM-KISTA) UNIFY ASSURA ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. (CRM-KISTA) UNIFY ASSURA ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3361-40Q
Device Problem Premature Discharge of Battery (1057)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/26/2015
Event Type  Injury  
Manufacturer Narrative
(b)(4) no conclusion code available.Final analysis found the reported premature battery depletion was confirmed in the laboratory.Based on all available parameter and usage information, device longevity was found to be below the expected limits.With an external power supply attached, the device functioned normally.No high current was detected during testing and the power consumption was normal.The original battery was returned to the vendor for further evaluation and analysis could not conclusively determine a root cause for the premature battery depletion.
 
Event Description
It was reported the device delivered a patient notifier for the device unexpectedly reaching eri.The device was explanted and replaced.The patient condition is fine.
 
Manufacturer Narrative
Device evaluation: premature battery depletion was confirmed by analysis.Bench testing on the device was performed, and no sources of high current were noted.In further analysis of the battery, lithium clusters were observed, but at the time of analysis the clusters did not appear to be in a location or size that would be sufficient to cause an internal short of the battery.As a result, the cause of the premature battery depletion could not be conclusively determined.However, from these analyses, in the absence of other root causes, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY ASSURA ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC. (CRM-KISTA)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC. (CRM-KISTA)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
ulla strindlund
box 7051
isafjordsgatan 15
kista SE-16-407
SW   SE-16407
4684744043
MDR Report Key5318454
MDR Text Key34125210
Report Number3010215456-2015-32665
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeSW
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date 06/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2015
Device Model NumberCD3361-40Q
Device Lot Number4085541
Other Device ID Number05414734508254
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/06/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/27/2015
Initial Date FDA Received12/22/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/05/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/20/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0116-2017
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-