Model Number CD3235-40
Device Problem Connection Problem (2900)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/05/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4) the reported capture anomaly was not confirmed in the laboratory.The device was tested on the bench and no anomalies were found.
 
Event Description
It was reported that during implant procedure increased lv thresholds were observed.Despite disconnecting and reconnecting the lead, the lv threshold remained high however, lead tested normal with the pacemaker system analyzer.The decision was made to use another device.Patient was stable post procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY CRT-D
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
jaime chavez
645 almanor avenue
sunnyvale, CA 94085
8184934022
MDR Report Key5321504
MDR Text Key34265572
Report Number2938836-2015-32273
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeIN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 10/05/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2016
Device Model NumberCD3235-40
Device Lot NumberS000004932
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/05/2015
Initial Date FDA Received12/22/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/14/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age46 YR
-
-