HILL-ROM BATESVILLE PROGRESSA BED; A/C POWERED ADJUSTABLE HOSPITAL BED Back to Search Results
Model Number 7500
Device Problem Device Alarm System (1012)
Patient Problem Hyperglycemia (1905)
Event Date 12/07/2015
Event Type  malfunction  
Manufacturer Narrative
The hill-rom technician found the external alarm defective.It is necessary for the progressa bed to have an effective maintenance program.We recommend that you do annual preventive maintenance.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the external alarm to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hill-rom received a report from a hill-rom technician stating the alarm on the device was not working.The bed was located at the account in 527.There was no patient/user injury reported.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRESSA BED
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
john cummings
1069 state route 46 east
batesville, IN 47006
8129312869
MDR Report Key5323306
MDR Text Key34767514
Report Number1824206-2015-01118
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 12/07/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other Caregivers
Device Model Number7500
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/07/2015
Initial Date FDA Received12/23/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/04/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-