• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECKMAN COULTER COULTER AC.T DIFF ANALYZER; COUNTER, DIFFERENTIAL CELL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECKMAN COULTER COULTER AC.T DIFF ANALYZER; COUNTER, DIFFERENTIAL CELL Back to Search Results
Model Number N/A
Device Problem Fluid/Blood Leak (1250)
Patient Problem Not Applicable (3189)
Event Date 12/07/2015
Event Type  malfunction  
Manufacturer Narrative
On (b)(6) 2015 the customer performed troubleshooting to fix the leak.The tubing in solenoids lv14 and lv15 was removed and inspected then placed back on by the customer.The instrument ran without leaks after that.Controls and startup passed after troubleshooting.(b)(4).
 
Event Description
The customer reported an uncontained clear leak of about 10-15 ml from the wbc (white blood cell) bath inside the act diff.Erroneous patient results were not generated and there was no change or effect to patient treatment in connection to the event.There was no impact to patient results and controls.The customer was wearing personal protective equipment (ppe) consisting of laboratory coat, goggles and gloves at the time of the event and there was no report of injury or biohazard exposure to open wounds or mucous membranes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COULTER AC.T DIFF ANALYZER
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 south kraemer blvd.
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th ave
miami 33196
Manufacturer Contact
miranda holland
11800 sw 147th ave
32-s08
miami 33196
3053802031
MDR Report Key5331713
MDR Text Key34575490
Report Number1061932-2015-01806
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K973634
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/07/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number6706366
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/07/2015
Initial Date FDA Received12/29/2015
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-